Back      Home      Remote Data Retrieval

Department of Chemistry
University of Toronto

CHM 410F/IES 1410F - Fall, 1998

Smoke II:
Field Sampling For ETS Derived Aldehydes And Nicotine

Field Sampling

This experiment will build on "Smoke I" by having each student go into the field to sample environmental tobacco smoke (ETS) with personal air monitoring pumps. To one of these you will attach 2,4 DNPH impregnated SPE cartridges and the other XAD-2 glass tube. The former will simultaneously derivitize and adsorb aldehydes as they are pulled through the cartridge while the latter will sorb nicotine. The targets of interest are nicotine, formaldehyde and any other aldehydes that can be resolved and detected. HPLC will be used to separate the formaldehyde hydrazone from other interfering compounds and capillary GC with nitrogen-phosphorous detection to selectively isolate and monitor nicotine.

Methodology

Air sampling. The nominal sampling rates for the two cartridges are 500 ml/min for the DNPH SPE cartridge and 1 liter/min for the XAD glass tube. You will need to calibrate the air pump, in the laboratory, both before and after you gather your sample by using a bubble meter. The TAs will demonstrate how to do this and will also give you some indication as to the total amount of air needed for the available detection limits attainable with the instruments in the lab; typically a sampling time of 1 hour at the above flow rates yields detectable signals.

Location. The purpose of this section VI is for you to carry out the sampling of polluted air at a location of your choice. Each student group will have to consider carefully the QA/QC issues involved in obtaining a representative sample, having appropriate blanks and controls, and maintaining sample integrity. The sample site should be chosen carefully to illustrate some hypothesis that is being made. For example if a smoke-filled bar is chosen for sampling perhaps the control site would be a smoke-free restaurant. Although tobacco smoke is likely the only source of nicotine it should be remembered that there are innumerable sources of formaldehyde (see references in Smoke I). Make sure to note as much about the sample location as possible for later data analysis.

Analysis. A lab period has been set aside for each group to extract the two analytes from their respective sorbents and measure each using the two chromatographic systems. The SPE cartridge will be eluted with acetonitrile (~5 ml) and diluted to the mark in a volumetric flask. The XAD tube will require a bit more work as there are two sections to the tube. The first and largest amount of XAD is the primary sorbent and you'll need to carefully remove this from the tube into an appropriate container containing the extracting solvent (ethyl acetate); be precise about the final volume of extracting solvent. The shorter section of XAD is the back-up sorbent and is there to indicate whether there was any "breakthrough" of nicotine during the air sampling. Please isolate this separately from the primary sorbent as it will represent a check on the sampling methodology. There, of course, should also have two DNPH cartridges attached to each other to provide the same assurance of no breakthrough but at this time we will not due to cost considerations. Upon obtaining each of the extracts proceed to the analysis step using the appropriate conditions/instruments that you are now familiar with.

Back      Home      Remote Data Retrieval