Cryogens
| Hazards of Cryogens Cryogenic liquids (or cryogens) are liquified gases that are cooled below room temperature-most cryogenic liquids are below -150°C. When a small amount of cryogenic liquid is converted into gas, a very large volume of gas is created. Under WHMIS, cryogenic liquids are classified as compressed gases. |
|
| Extreme Cold Cryogens can freeze flesh, causing painful blisters, much like a burn. Prolonged exposure can cause frostbite with pain occuring only when the flesh thaws. Flesh can stick to cold metals. |
|
| Asphyxiation Cryogens expand into large volumes of gas that can displace air. For example, 1L of liquid nitrogen forms nearly 700L of nitrogen gas at room temperature. The gas formed is often cold and pools on the floor or lower areas. In enclosed areas, death or coma from oxygen deficiency may occur. Do not enter an oxygen deficient atmosphere even to rescue someone. Always store dewars in well-ventilated areas. Never enter the cryogen facility if the oxygen warning sensor alarm is sounding. The oxygen level alarm and sensor are located on the wall next to the freight elevator in the Cryogenic Facility. |
|
| Toxic Hazards Toxic cryogens will release toxic gases. Read the MSDS that came with the liquid. |
|
| Obscured Vision The vapour formed from cryogens fall down forming a ground level fog that obscures the floor. Beware of trip hazards. |
|
| High Pressure Sealed systems containing cryogens may form extremely high pressures, enough to rupture or explode. Always have a relief vent on a cryogen dewar. |
|
| Dewars in High Magnetic Fields Superconducting magnets are rountinely filled with cryogens. The dewars used for this purpose must be non-magnetic. |
|
| Liquid Oxygen Liquid oxygen can make materials burn that are considered non-combustable like teflon or aluminum. Organic materials may react explosively. Liquid oxygen often forms on the outside of transfer lines. |
DO | DO NOT |
|
|
|
Alarm levels in the Cryogenic Facility
| Oxygen level | Alarm |
| 19.5% | Intermittent alarm Continous alarm and solenoid valves shut off the N2 system. Leave the area at once and report this to Ken Greaves in LM20. |
| 19.1% | The oxygen level alarm and sensor are located on the wall next to the freight elevator in the Cryogenic Facility. |